Standard Full Field White on White Perimetry The IVS series offer a wide range of test patterns and strategies, including T30-2, T24-2, T10-2 for glaucoma diagnosis and T-Macula for macula function assessment. In addition, special test patterns like driver feasibility, monocular/binocular social security disability are also included. For better detection of visual field loss caused by early stage glaucoma, points of T30-2 and T24-2 are cautiously configured on the most sensitive position of retinal nerve fiber bundle. |
Precise Diagnosis Strictly conforming to the newest perimetry standard of IMAGING and PERIMETRY SOCIETY, IVS series simultaneously fulfill the needs for ophthalmological and neurological uses. Equipped with world-wide accepted 31.5 asb background illumination and incorporated with efficient HISA algorithm, comprehensive fixation control and age-related normal database, IVS’s test result is highly consistent with Goldmann standard perimeter. Its aspheric dome effectively shrinks its size, while ensuring the full field test range. | Low Light-declining LED No Need for Replacement & Calibration LED light source has characteristics of high efficiency, low heat, long life and low light decay. Its radiating energy reduction is less than 20% after 30,000 hours of illumination. That means, if there is about 300 stimulus for each test, and each stimuli lasts for 200 ms, then, after 1.8 million tests, the stimulus intensity of IVS has attenuated less than 1 dB. Furthermore, to ensure even more stable performance, IVS has built a light measuring sensor internally to detect any abnormal change in light intensity. | |
Gaze Tracking Minimize effects of unreliable response “Gaze Tracking” constantly monitors the pupil position and patient’s fixation. System beeps to draw patient’s attention when a fixation shift is detected. If fixation shift lasts for a while, system will stop test and inquire operator’s intervention. | Blink Control Never miss a stimuli IVS’s BLINK CONTROL helps patients avoid from dry eyes and relax them to stay focused in subsequent test. When IVS’s BLINK CONTROL is on, stimuli during patient’s blink will not be counted and will automatically be repeated in later test. | |
Dedicated System, Safer Data, Easier Use Instead of common x86 PC and its relative operating system, IVS series adopt low power, high reliability, space compact industrial embedded ARM computer and dedicated operating system. IVS series is immune to common computer viruses, and its simple structure (no video card, no hard disk, no DVD drive) brings incredible reliability. More 1,000,000 tests can be stored in more than 32 GB solid state disk. High drop and unusual power off will not result in any data loss. | HISA-Heuristic Interactive Threshold Searching Algorithm HISA forecasts initial threshold for new point through a complex mathematical model, which takes neighboring tested results and same age normal values into consideration. Then unnecessary search will consequently be avoided. HISA does a sample survey from some specific location. Subsequently, time will be saved for patients with serious reduction of sensitivity by skipping those high-sensitivity questions. | |
Glaucoma Progression Analysis IVS can accurately differentiate clinically significant progression of visual field loss from random variability within a series of follow up tests, providing an advanced, reliable method to enhance the management of glaucoma. It really helps identify rapidly progressing, high-risk patients. | Automated Pupil Measurement IVS can automatically measure patient’s pupil diameter and print it in report. This benefits comprehension of the correlation between pupil size and perimetry result to avoid wrong report interpretation with too small pupil size. | |
Powered by Froala Editor
Model No. | IVS-201B | IVS-201A | |||
---|---|---|---|---|---|
Stimulus Generation | Hidden LED Array | ||||
Max Temporal Range | 90° | ||||
Testing Distance | 30cm | ||||
Background Illumination | 31.5asb (10cd/m2) | 315asb (100cd/m2) | |||
Stimulus Size | Goldmann III | ||||
Stimulus Intensity | 0.08asb ~ 10,000asb (0 ~ 50 dB) | ||||
Stimulus Duration | 200ms, configurable | ||||
Test Strategy | |||||
Threshold Test Library | T10-2, T24-2, T30-2, T-Macula, T60-4 | ||||
Threshold Strategy | Full Threshold, Fast Ladder, HISA, Standard Threshold | ||||
Screening Test Library | S-64, S-76 | S-40, S-64, S-60,S-76 | |||
Screen Strategy | Two Zone, Three Zone, Qualified Defect | ||||
Blue/Yellow Perimetry | N/A | T24-SWAP, T30-SWAP | |||
Custom Testing | N/A | Customized Program | |||
Colored Perimetry | N/A | Blue Stimuli | |||
Specialty Test Patterns | D-30, D-60, EM-M, EM-B | ||||
Fixation Monitor | Heijl/Krakau blind spot monitor, Infrared video eye monitor, Gaze tracking, Fixation tracking, Pupil measurement, Blink control | ||||
Software Features | Visual Field Index, Glaucoma Hemifield Test(GHT), Single field analysis,Serial field overview, HISA Analysis Available in IVS-201A only: SWAP analysis, Networking, Glaucoma Progression Analysis, DICOM | ||||
Responser | Hand held, Foot pedal (optional) | ||||
Built-in System | |||||
Operating System | Dedicated OS (immune for general computer viruses) | ||||
Operator Interface | 15’’ LCD touch screen, Keyboard & Mouse(option) | ||||
Data Storage | ≥32GB, More than 1,000,000 test results | ||||
Data Backup | Flash Disk, Networking | ||||
Networking | Ethernet |